Fullbrook Systems
Fullbrook Systems - © 2011 #2 Contact us.
Kemtak home Kemtrak NBP007 size & concentration Kemtrak TCP007 turbidity measurement DCP007 Colour and concentration DCP007 NIR Kemtrak Applications Data logging

TC007 Industrial Turbidimeter

 

The Kemtrak TC007 turbidimeter is a robust in-line turbidimeter designed for a wide range of industrial process applications. The Kemtrak TC007 uses the nephelometric method of turbidity measurement in accordance with International Standard ISO7027:1999(E).

 

The Kemtrak TC007 has an easy to use menu based software with a user configurable display that shows the turbidity value in real time. Additionally, the TC007 has a database and event handler to capture vital process information for use in quality assurance and plant control reports. All data is easily transferable to a PC via a USB interface.

 

Optic fibers are used to pipe the light from within the TC007 enclosure to the sampling point and back. This is a major benefit as all the electricity is kept within the analyzer enclosure permitting safe operation in even the most hazardous of environments.

 

Optic fibres can be ordered in lengths up to 50m and are well suited for harsh industrial environments. Our standard fibers have a silicone-coated steel jacketing and can withstand temperatures up to 200 °C. A range of robust flow-cell manifolds are also available dependent upon the process and application.

Main features:

 

   * Long life 880nm LED light source

   * Simple & robust flowcell design using fibre

       optics - no electricity or heat near the sample

   * Easy to operate - four push buttons, user-

       friendly menu and alphanumeric display

   * Two alarms (high/low) and system fault

        monitoring

   * Cleaning sequence control (relay output)

   * PID Controller (pulse width modulated relay or

     4-20 mA analogue output)

   * Analogue output (0-20/4-20 mA)

   * Data & system event logger for quality control

   * PC downloading software (Windows 2000,        XP & Vista)

   * Supply voltage 110/240 VAC

   * IP65 protection

   * Overall dimensions:220 x 120 x 100 mm        (LxWxD)

 

* wavelengths must be factory installed

Sensor flange example - DIN flange DN25